Prediction of the Prosthetic Valve size by use of Supraaortic Cineangiogram

Cineangiogram 을 이용한 대동맥판막의 수술전 인공판막 치수의 예

  • Lee, Young-Thak (Department of Thoracic Surgery, Seoul National University Hospital) ;
  • Ahn, Hyuk (Department of Thoracic Surgery, Seoul National University Hospital) ;
  • Park, Jae-Hyung (Department of Diagnostic Radiology, Seoul National University Hospital)
  • 이영탁 (서울대학병원 흉부외과) ;
  • 안혁 (서울대학병원 흉부외과) ;
  • 박재형 (서울대학병원 방사선과)
  • Published : 1987.03.01

Abstract

We compared between prosthetic aortic valve size and aortic annulus size in supravalvular aortic cineangiogram in 30` RAO view postoperatively. Retrospectively, supraaortic cineangiogram of 27 patients among the patients underwent aortic valve replacement only or double valve replacement from April, 1986 to January, 1987 was examined and measured the aortic annulus size. In comparing the two values, the cases within 1mm is 22, and the cases within 2mm is 25, correlation coefficient yield r = 0.92. In two cases, the difference between two values is within 3mm We concluded that to prevent the complication from mismatching the prosthetic aortic valve size to patient`s annulus size [e.g. left ventricular failure, hemolysis, limited exercise tolerance], prediction of the prosthetic valve size preoperatively by use of cineangiogram is useful.