[ $PERIOTEST^{(R)}$ ] ASSESSMENT OF DENTAL IMPLANT MOBILITY : A CLINCAL STUDY

$PerioTest^{(R)}$를 이용한 임플란트 동요도에 관한 임상적 연구

  • Kim, Sun-Jai (Department of Dentistry, The graduate School, Yonsei University) ;
  • Han, Dong-Hoo (Department of Dentistry, The graduate School, Yonsei University)
  • 김선재 (연세대학교 치과대학 치과보철학교실) ;
  • 한동후 (연세대학교 치과대학 치과보철학교실)
  • Published : 1998.10.01

Abstract

Assessment of implant mobility is an important and reliable method to clinically evaluate implant stability. PerioTest is a precise and reproducible device that cam dynamically measure the reaction of damping characteristics of peri-implant tissue. The aim of this study is to evaluate the effects of amount of implant surface area, diameter, type, implantation site, degrees of cortical engagement, and length of time in function on PTVs and to find out the most determining factor on PTVs. The results are as follows 1. 5.0mm diameter implants show significantly lower PTV than that of 3.75mm diameter implants. 2. PTV in the mandible is significantly lower than that of the maxilla 3. In the maxilla, there is no significant difference in PTV during the first year of implant function, but during the second year a significant decrease in PTV is noted. 4. In the mandible, there is a significant decrease in PTV during the first and second year of implant function. 5. Implantation site seems to be the most determining factor on PTV among the influencing factors in this study. In conclusion, the amount of implant surface area, type, degrees of cortical engagement had no significant effect on PTV, but installation site and diameter influenced significantly on PTV

Keywords