Characteristic Feature of Inductively Coupled Plasma Atomic Emission Spectrometer/Shielding System and Evaluation of Its Applicability to Analysis of Radioactive Materials

유도 결합 플라스마 원자방출분광기/차폐 시스템의 특성 및 방사성 물질 분석에 대한 적용성 평가

  • Lee, Chang Heon (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute) ;
  • Suh, Moo Yul (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute) ;
  • Choi, Kae Chun (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute) ;
  • Park, Yang Soon (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute) ;
  • Jee, Kwang Yong (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute) ;
  • Kim, Won Ho (Nuclear Chemistry Research Team, Korea Atomic Energy Research Institute)
  • 이창헌 (한국원자력연구소 원자력화학연구팀) ;
  • 서무열 (한국원자력연구소 원자력화학연구팀) ;
  • 최계천 (한국원자력연구소 원자력화학연구팀) ;
  • 박양순 (한국원자력연구소 원자력화학연구팀) ;
  • 지광용 (한국원자력연구소 원자력화학연구팀) ;
  • 김원호 (한국원자력연구소 원자력화학연구팀)
  • Received : 2000.04.29
  • Published : 2000.08.25

Abstract

An inductively coupled plasma atomic emission spectrometer/shielding system was specially designed and built for the analysis of radioactive materials. Both of an inductively coupled plasma source and a sample transfer system to be contacted with radioactive materials was installed in a stainless steel glove box. In terms of analytical capability and radiation safety, characteristic feature of the system was investigated. Its applicability to the determination of fission products and corrosion products in the radioactive materials such as spent fuel dissolver solution and the primary coolant of nuclear power reactors was evaluated. In the concentration range $0.01-0.1mgL^{-1}$, the relative standard deviation was found to be less than 5%.

사용후핵연료에 함유되어 있는 핵분열생성물을 분석하기 위하여 여러 원소를 동시에 분석할 수 있고, 분석감도가 커서 시료의 방사능과 폐기물의 양을 감소시킬 수 있는 유도 결합 플라스마 원자방출분광기/차폐 시스템을 구성하였다. 방사성 물질이 직접 접촉되는 플라스마 들뜸원과 시료용액 도입부를 스테인리스 스틸 재질의 글로브박스 내부에 설치하였으며, 고주파 들뜸전원, 분광기, 검출기 그리고 전기, 전자 및 아르곤 가스 공급 제어장치는 외부에 설치하였다. 분석능과 방사선 안전의 관점에서 시스템의 특성을 검증하였으며, 사용후핵연료 용해용액과 원자력발전소의 일차냉각수를 대상으로 핵분열생성물과 방사성 부식생성물 분석에 대한 적용성을 평가한 결과 $0.01-0.1mgL^{-1}$ 농도범위에서 상대표준편차는 5% 이하였다.

Keywords

References

  1. Atomic Spectrosc v.7 K. Kato
  2. PNL-5862 Shielded analytical instruments for characterization of highly radioactive materials J. L. Daniel;H. E. Kjarmo
  3. LA-UR-86-889 The Los Alamos installation for the inductively coupled plasma atomic emission spectrometric analysis of radioactive solutions C. T. Apel
  4. IS-4557 Optical assay technology for safeguards M. C. Edelson;V. A. Fassel
  5. IS-4774 An extended work plan the simultaneous assay uranium and plutonium, and their isotopes, by optical emission spectroscopy M. C. Edelson;V. A. Fassel
  6. IS-M 336 Application of inductively coupled plasma emission and fluorescence spectroscopy to the safeguarding of a nuclear facility M. C. Edelson;V. A. Fassel
  7. IS-M 337 Application of ICP-AES to the multi-elemental and multi-isotopic analysis of fuel dissolver solution M. C. Edelson;V. A. Fassel
  8. IS-4800, UC-15 Optical spectrometric SNM assay M. C. Edelson;V. A. Fassel
  9. J. Plasma, Spectro. v.4 K. Onishi;T. Siganuma
  10. JAERI-M 90-062 Design, construction and performance test of atomic emission spectrography with DCA and ICP excitation system for plutonium bearing fuels S. Fukushima;M. Handa;K. Shiozawa;M. Hirata;H. Mizukami
  11. ASTM C 852-81 Standard for design criteria for plutonium gloveboxes
  12. ASTM C 852-93 Standard for design criteria for plutonium gloveboxes
  13. 과학기술부고시 제98호-12호 4조
  14. Grove Engineering
  15. Annual report Chang Heon Lee
  16. Analytical Chemistry Symposia Series v.19 Consideration for the installation of an inductively coupled plasma for the analysis of radioactive samples C. T. Apel;D. L. Gallimore
  17. Anal. Chem. v.64 Eric H. van Veen;Margaretha T. C. de loos-Vollebregt
  18. Spectrochim. Acta v.42B E. A. Huff
  19. Appl. Spectrosc. v.43 E. A. Huff;D. L. Bowers
  20. LA Report No. 4400 Los Alamos Handbook of Radiation Monitoring J. W. Healy
  21. Analytical Chemistry Symposia Series v.19 A new facility for high resolution inductively coupled plasma-atomic multielement emission spectroscopy of radioactive materials M. C. Edelson;V. A. Fassel
  22. NUREG/CP-0098, Vol. 1;CONF-880815 v.1 Efficiency of HEPA filters at elevatedtemperatures;investigations with the TiO₂ test method U. Ensinger;V. Ruedinger;J. G. Wilhelm;M. W. First
  23. NUREG/CP-0130, Vol. 1; CONF-920823 v.1 The effect of temperature on HEPA filter media C. Hambline;P. J. Goodchild
  24. Spectrochim. Acta v.41B M. C. Edelson;E. L. DeKalb;R. K. Winge;V. A. Fassel