Measurement of a Blood Velocity by using Photoplethysmograph and Radial Artery Pulse Wave Equipped with Magnetic Hall Device

자성 홀소자 맥진기와 용적맥파계의 맥진파형을 이용한 혈류속도 측정 연구

  • Jang, Deok-Hyeong (Mediana Co., Ltd., Dongwha Medical Instrument Complex) ;
  • Kim, Dam-Bee (Dept. of Oriental Biomedical Engineering, Sangji University) ;
  • Choi, Suel-Gi (Dept. of Oriental Biomedical Engineering, Sangji University) ;
  • Lee, Sang-Suk (Dept. of Oriental Biomedical Engineering, Sangji University)
  • Received : 2012.07.09
  • Accepted : 2012.08.01
  • Published : 2012.08.31

Abstract

One prototype product of clip-type pulsimeter equipped with magnetic field sensing semiconductor Hall device after one permanent magnet attached "Chwan" position in center of a radial artery was developed. The clip-pulsimeter was composed of the hard ware system measuring to voltage signals. To measure the blood velocity, the radial artery pulsimeter is simultaneously connected the PPG (photoplethysmograph). Analysis and comparison of two pulse waves data has done obtained from a clinical test of forty subjects of 20 ages. The value of a blood velocity simultaneously measured from a radial artery puls wave and PPG is an average value of 0.8m/s. The usage of this research results is possible to store the biomedical signals for health care.

요골동맥의 중심인 "관" 위치에 영구자석을 붙인 후 바로 위쪽에 자성측정용 반도체 홀소자를 장착하여 전압 신호를 검출하는 하드웨어 시스템이 적용된 집게형 맥진기 시제품을 개발하였다. 집게형 맥진기와 용적맥파계를 동시 분석하는 시스템으로 20대 나이 임상참가자 40명 각각에 대한 맥진파형과 용적맥파형 두 피크값을 비교 측정한 결과 혈류속도가 평균 0.8 m/s로 기존 값과 거의 비슷하였다. 이로써 요골동맥파형-용적맥파형을 환자감시장치에서 제시하는 주요 임상의 주요값으로 적용할 수 있음을 제시하였다.

Keywords

References

  1. S. S. Lee, D. H. Nam, Y. S. Hong, W. B. Lee, I. H. Son, K. H. Kim, and J. G. Choi, Sensors 11, 1784 (2011). https://doi.org/10.3390/s110201784
  2. P. A. Shaltis, A. T. Reisner, and H. H. Asada, IEEE Trans. Biomed Eng. 55, 1775 (2008). https://doi.org/10.1109/TBME.2008.919142
  3. M. C. Ahn, J. G. Choi, I. H. Son, S. S. Lee, and K. H. Kim, J. Kor. Mag. Soc. 20, 106 (2010). https://doi.org/10.4283/JKMS.2010.20.3.106
  4. S.-S. Lee, I.-H. Son, J.-G. Choi, D.-H. Nam, Y.-S. Hong, and W.-B. Lee, J. Kor. Phys. Soc. 58, 349 (2011). https://doi.org/10.3938/jkps.58.349
  5. I.-H. Son and S.-S. Lee, J. Kor. Mag. Soc. 21, 104 (2011). https://doi.org/10.4283/JKMS.2011.21.3.104
  6. M. F. P. O'Rourke, R. P. Kelly, and A. P. Avolio, The Arterial Pulse, 1st Ed.; Lea & Febiger, Philadelphia, PA, USA (1992).
  7. S. S. Lee, M. C. Ahn, and S. H. Ahn, J. Magnetics 14, 132 (2009). https://doi.org/10.4283/JMAG.2009.14.3.132
  8. John Allen, Photoplethysmography and its Application in Clinical Physiological Measurement 28, R1 (2007).
  9. K. Shelley and S. Shelley, Pulse Oximeter Waveform: Olethysmography, in Clinical Monitorning, Carol Lake, R.Hines, and C. Blitt, Eds.: W. B. Saunders Company (2001) pp 420-428.
  10. A. Fronek, Noninvasive diagnosis in vascular disease, McGraw Hill, New York (1989).
  11. R. Asmar, A. Benetos, G. M. London, C. Hughe, Y. Weiss, J. Topouchian, et al., Blood Pressure 4, 48 (1995). https://doi.org/10.3109/08037059509077567
  12. Q. Yu, J. Zhou, and Y. C. Fung, Am. J. Physiol. Heart Circ. Physiol. 265, 52 (1993).

Cited by

  1. Extraction of Respiratory Rate by using FFT for Radial Artery Pulse Waves Acquisited by Clip-type Pulsimeter with a Hall Sensor vol.22, pp.5, 2012, https://doi.org/10.4283/JKMS.2012.22.5.178
  2. Development of Oriental-Western Fusion Patient Monitor by Using the Clip-type Pulsimeter Equipped with a Hall Sensor, the Electrocardiograph, and the Photoplethysmograph vol.23, pp.4, 2013, https://doi.org/10.4283/JKMS.2013.23.4.135