DOI QR코드

DOI QR Code

Evaluation for Performance of a Military Radiation Detecter PDR-1K : Focused on Dependence of Angular and Energy

군사용 방사선 계측기 PDR-1K 반응도 평가 : 방향 및 에너지 의존성을 중심으로

  • Park, Wonseok (The 85 Precision Standard Maintenance Depot, Air Force) ;
  • Choi, Junhyuk (Department of Emergency Management, Inje University) ;
  • Jung, Doyoung (Department of Emergency Management, Inje University) ;
  • Kim, Jango (Department of Emergency Management, Inje University) ;
  • Min, Byungin (Department of Emergency Management, Inje University)
  • 박원석 (공군 85정밀표준정비창) ;
  • 최준혁 (인제대학교 재난관리학과) ;
  • 정도영 (인제대학교 재난관리학과) ;
  • 김장오 (인제대학교 재난관리학과) ;
  • 민병인 (인제대학교 재난관리학과)
  • Received : 2016.10.11
  • Accepted : 2017.03.24
  • Published : 2017.04.05

Abstract

In this paper, we performed a evaluation for angular and energy dependence of military radiation detector PDR-1K. Its measuring range is divided into two section, low and high, and each range has a GM tube separately owing to broad scale. We observed a change in relative angular reactivity within 0.928 ~ 1.188 in low range and within 0.743 ~ 1.000 in high range from $-90^{\circ}$ to $+90^{\circ}$. The evaluation for energy dependence appeared a relative reactivity within 0.892 ~ 1.000 above 83 keV. This means PDR-1K isn't possible to use below 83 keV of radiation energy. It is possible to provide response information to user and to increase reliability of radiation measurement through this paper.

Keywords

References

  1. Choi, G. W. et al., "Development of a Techniques of the Performance Test for a Radiation Protection Devices and it's International Standards," Journal of Radiation Protection and Research, Vol. 33, No. 1, pp. 1-12, 2008.
  2. Glenn F. Knoll, "Radiation Detection and Measurement," John Wiley & Sons, Inc. Newyork, pp. 201-216, 1999.
  3. Choi, G. W. et al., "Development of Characteristic Evaluation Technics for Radiation Detectors," KRISS, Korea, pp. 1-3, 2005.
  4. KATS, "Radiation Protection Instrumentation-Ambient and/or Directional Dose Equivalent(Rate) Meters and/or Monitors for Beta, X and Gamma Radiation-Part 1 : Portable Workplace and Environmental Meters and Monitors," KS C IEC 60846-1, Korea, 2014.
  5. Korea Association of Standards & Testing Organization (KASTO), "Standard Calibration Procedure of X and Gamma Radiation Survey Meters," KASTO 02-26-4030-034, Korea, 2002.
  6. International Organization for Standardization(ISO), "X and Gamma Reference Radiation for Calibrating Dosemeters and Doserate Meters and for Determining Their Response as a Function of Photon Energy-Part 1: Radiation Characteristics and Production Methods," ISO 4037-1, Geneva, 1996.
  7. Ha, S. H., "Evaluation for Stanadard Gamma Irradiator Performence of 85 PSMD," KRISS, Korea, pp. 1-9, 2015.
  8. International Electronical Committee(IEC), "Radiation Protection Instrumentation Measurement of Personal Dose Equivalents Hp(10) and Hp(0.07) for x, Gamma, Neutron and Beta Radiations-Direct Reading Personal Dose Equivalent Meters and Monitors and Personal Warning Devices," IEC61526, Geneva, 2002.