A Commissioning Report on the Magnetic Resonance-Compatible Geneva Brachytherapy Applicator

  • Yoonsuk Huh (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Hyojun Park (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jin Jegal (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Inbum Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jaeman Son (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Seonghee Kang (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Chang Heon Choi (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jung-in Kim (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Hyeongmin Jin (Department of Radiation Oncology, Seoul National University Hospital)
  • Received : 2024.09.30
  • Accepted : 2024.12.11
  • Published : 2024.12.31

Abstract

Purpose: Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning further promoted the adoption of advanced techniques such as the Geneva applicator. This study aims to commission the Geneva applicator, focusing on wall thickness, dummy marker positions, and source dwell positions to ensure accurate dose delivery and safety. Methods: The commissioning process involved measuring wall thickness in both the longitudinal and transverse directions for the tandem and lunar-shaped ovoid tubes and comparing these measurements with the manufacturer's specifications. Dummy marker positions were verified using CT imaging, with a focus on alignment tolerances of ±1 mm. Source dwell positions were planned using the Oncentra treatment planning system, with measurements taken using EBT4 film and analyzed with RIT software. Results: Wall thickness measurements and dummy marker positions were within the specified tolerance ranges, confirming their accuracy. The source dwell positions, measured and analyzed through multiple tests, were all within the ±1 mm tolerance, ensuring the applicator's reliability. Conclusions: The Geneva applicator met all standards for safe and effective use in brachytherapy. The use of a 3D-printed holder was crucial for precise alignment and measurement. With updated reimbursement policies in Korea for MRI-based brachytherapy, the Geneva applicator is expected to significantly impact the future of advanced brachytherapy treatments and research.

Keywords

Acknowledgement

This study was supported by a grant from the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP) (No. 2022R1F1A1063779) and a grant No. 0720233109 from the SNUH Research fund.

References

  1. Glaser SM, Mohindra P, Mahantshetty U, Beriwal S. Complications of intracavitary brachytherapy for gynecologic cancers and their management: a comprehensive review. Brachytherapy. 2021;20:984-994. https://doi.org/10.1016/j.brachy.2020.11.011
  2. Tanderup K, Eifel PJ, Yashar CM, Pötter R, Grigsby PW. Curative radiation therapy for locally advanced cervical cancer: brachytherapy is NOT optional. Int J Radiat Oncol Biol Phys. 2014;88:537-539. https://doi.org/10.1016/j.ijrobp.2013.11.011
  3. Brady LW, Micaily B, Miyamoto CT, Heilmann HP, Montemaggi P. Innovations in brachytherapy in gynecologic oncology. Cancer. 1995;76(10 Suppl):2143-2151. https://doi.org/10.1002/1097-0142(19951115)76:10+<2143::AID-CNCR2820761339>3.0.CO;2-4
  4. Tanderup K, Viswanathan AN, Kirisits C, Frank SJ. Magnetic resonance image guided brachytherapy. Semin Radiat Oncol. 2014;24:181-191. https://doi.org/10.1016/j.semradonc.2014.02.007
  5. Schoeppel SL, Ellis JH, LaVigne ML, Schea RA, Roberts JA. Magnetic resonance imaging during intracavitary gynecologic brachytherapy. Int J Radiat Oncol Biol Phys. 1992;23:169-174. https://doi.org/10.1016/0360-3016(92)90557-X
  6. Zhang H, Donnelly ED, Strauss JB, Kang Z, Gopalakrishnan M, Lee PC, et al. Clinical implementation, logistics and workflow guide for MRI image based interstitial HDR brachytherapy for gynecological cancers. J Appl Clin Med Phys. 2019;20:37-49. https://doi.org/10.1002/acm2.12736
  7. Dolezel M, Odrazka K, Vanasek J, Kohlova T, Kroulik T, Kudelka K, et al. MRI-based pre-planning in patients with cervical cancer treated with three-dimensional brachytherapy. Br J Radiol. 2011;84:850-856. https://doi.org/10.1259/bjr/75446993
  8. Kirisits C, Pötter R, Lang S, Dimopoulos J, Wachter-Gerstner N, Georg D. Dose and volume parameters for MRI-based treatment planning in intracavitary brachytherapy for cervical cancer. Int J Radiat Oncol Biol Phys. 2005;62:901-911. https://doi.org/10.1016/j.ijrobp.2005.02.040
  9. Hellebust TP. Place of modern imaging in brachytherapy planning. Cancer Radiother. 2018;22:326-333. https://doi.org/10.1016/j.canrad.2018.03.005
  10. Song WY, Tanderup K, Pieters B. Emerging technologies in brachytherapy. CRC Press; 2017.
  11. Mour ya A, Aggarwal LM, Choudhar y S. Evolution of brachytherapy applicators for the treatment of cervical cancer. J Med Phys. 2021;46:231-243. https://doi.org/10.4103/jmp.jmp_62_21
  12. Rivard MJ, Coursey BM, DeWerd LA, Hanson WF, Huq MS, Ibbott GS, et al. Update of AAPM Task Group No. 43 Report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys. 2004;31:633-674. https://doi.org/10.1118/1.1646040