Development of an Instantaneously Interpretable Real-Time Dosimeter System for Quality Assurance of a Medical Linear Accelerator

  • Dongyeon Lee (Department of Radiation Convergence Engineering, Yonsei University) ;
  • Sung Jin Kim (Department of Radiation Oncology, Samsung Medical Center) ;
  • Wonjoong Cheon (Department of Radiation Oncology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Hyosung Cho (Department of Radiation Convergence Engineering, Yonsei University) ;
  • Youngyih Han (Department of Radiation Oncology, Samsung Medical Center)
  • Received : 2024.09.23
  • Accepted : 2024.12.13
  • Published : 2024.12.31

Abstract

Purpose: Modern radiotherapy delivers radiation doses to targets within a few minutes using intricate multiple-beam segments determined with multi-leaf collimators (MLC). Therefore, we propose a scintillator-based dosimetry system capable of assessing the dosimetric and mechanical performance of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) in real time. Methods: The dosimeter was equipped with a scintillator plate and two digital cameras. The dose distribution was generated by applying deep learning-based signal processing to correct the intrinsic characteristics of the camera sensor and a tomographic image reconstruction technique to rectify the geometric distortion of the recorded video. Dosimetric evaluations were performed using a gamma analysis against a two-dimensional array and radiochromic film measurements for 20 clinical cases. The average difference in the MLC position measurements and machine log files was tested for the applicability of the mechanical quality assurance (QA) of MLCs. Results: The agreement of the dose distribution in the IMRT and VMAT plans was clinically acceptable between the proposed system and conventional dosimeters. The average differences in the MLC positions for the IMRT/VMAT plans were 1.7010/2.8107 mm and 1.4722/2.7713 mm in banks A and B, respectively. Conclusions: In this study, we developed an instantaneously interpretable real-time dosimeter for QA in a medical linear accelerator using a scintillator plate and digital cameras. The feasibility of the proposed system was investigated using dosimetric and mechanical evaluations in the IMRT and VMAT plans. The developed system has clinically acceptable accuracy in both the dosimetric and mechanical QAs of the IMRT and VMAT plans.

Keywords

Acknowledgement

This work was supported by Korea Institute for Advancement of Technology (KIAT) grant funded by the Korea Government (MOTIE) (P0012971).

References

  1. Brahme A. Optimization of stationary and moving beam radiation therapy techniques. Radiother Oncol. 1988;12:129-140. https://doi.org/10.1016/0167-8140(88)90167-3
  2. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317. https://doi.org/10.1118/1.2818738
  3. Ling CC, Burman C, Chui CS, Kutcher GJ, Leibel SA, LoSasso T, et al. Conformal radiation treatment of prostate cancer using inversely-planned intensity-modulated photon beams produced with dynamic multileaf collimation. Int J Radiat Oncol Biol Phys. 1996;35:721-730. https://doi.org/10.1016/0360-3016(96)00174-5
  4. Bortfeld T, Boyer AL, Schlegel W, Kahler DL, Waldron TJ. Realization and verification of three-dimensional conformal radiotherapy with modulated fields. Int J Radiat Oncol Biol Phys. 1994;30:899-908. https://doi.org/10.1016/0360-3016(94)90366-2
  5. Chui CS, LoSasso T, Spirou S. Dose calculation for photon beams with intensity modulation generated by dynamic jaw or multileaf collimations. Med Phys. 1994;21:1237-1244. https://doi.org/10.1118/1.597206
  6. Van Esch A, Bohsung J, Sorvari P, Tenhunen M, Paiusco M, Iori M, et al. Acceptance tests and quality control (QC) procedures for the clinical implementation of intensity modulated radiotherapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments. Radiother Oncol. 2002;65:53-70. https://doi.org/10.1016/S0167-8140(02)00174-3
  7. Low DA, Moran JM, Dempsey JF, Dong L, Oldham M. Dosimetry tools and techniques for IMRT. Med Phys. 2011;38:1313-1338. https://doi.org/10.1118/1.3514120
  8. Han Y. Review on the pre-treatment quality assurance for intensity modulated radiation therapy. Prog Med Phys. 2013;24:213-219. https://doi.org/10.14316/pmp.2013.24.4.213
  9. Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36:5359-5373. https://doi.org/10.1118/1.3238104
  10. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al.; Task Group 142, American Association of Physicists in Medicine. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197-4212. https://doi.org/10.1118/1.3190392
  11. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: recommendations of AAPM Task Group No. 218. Med Phys. 2018;45:e53-e83. https://doi.org/10.1002/mp.12810
  12. Mehrens H, Taylor P, Followill DS, Kry SF. Survey results of 3D-CRT and IMRT quality assurance practice. J Appl Clin Med Phys. 2020;21:70-76. https://doi.org/10.1002/acm2.12885
  13. Ju SG, Han Y, Kum O, Cheong KH, Shin EH, Shin JS, et al. Comparison of film dosimetry techniques used for quality assurance of intensity modulated radiation therapy. Med Phys. 2010;37:2925-2933. https://doi.org/10.1118/1.3395574
  14. Sharma M, Singh R, Dutt S, Tomar P, Trivedi G, Robert N. Effect of absorbed dose on post-irradiation coloration and interpretation of polymerization reaction in the Gafchromic EBT3 film. Radiation Physics Chem. 2021;187:109569. https://doi.org/10.1016/j.radphyschem.2021.109569
  15. McCurdy BM, Luchka K, Pistorius S. Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device. Med Phys. 2001;28:911-924. https://doi.org/10.1118/1.1374244
  16. Van Esch A, Depuydt T, Huyskens DP. The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields. Radiother Oncol. 2004;71:223-234. https://doi.org/10.1016/j.radonc.2004.02.018
  17. Lee C, Menk F, Cadman P, Greer PB. A simple approach to using an amorphous silicon EPID to verify IMRT planar dose maps. Med Phys. 2009;36:984-992. https://doi.org/10.1118/1.3075817
  18. Warkentin B, Steciw S, Rathee S, Fallone BG. Dosimetric IMRT verification with a flat-panel EPID. Med Phys. 2003;30:3143-3155. https://doi.org/10.1118/1.1625440
  19. McCurdy BM, Greer PB. Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. Med Phys. 2009;36:3028-3039. https://doi.org/10.1118/1.3148822
  20. Nelms BE, Rasmussen KH, Tome WA. Evaluation of a fast method of EPID-based dosimetry for intensity-modulated radiation therapy. J Appl Clin Med Phys. 2010;11:3185. https://doi.org/10.1120/jacmp.v11i2.3185
  21. Ansbacher W. Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans. Med Phys. 2006;33:3369-3382. https://doi.org/10.1118/1.2241997
  22. Rowshanfarzad P, Sabet M, O'Connor DJ, Greer PB. Improvement of Varian a-Si EPID dosimetry measurements using a lead-shielded support-arm. Med Dosim. 2012;37:145-151. https://doi.org/10.1016/j.meddos.2011.06.003
  23. Ko L, Kim JO, Siebers JV. Investigation of the optimal backscatter for an aSi electronic portal imaging device. Phys Med Biol. 2004;49:1723-1738. https://doi.org/10.1088/0031-9155/49/9/010
  24. Rowshanfarzad P, Sabet M, O'Connor DJ, Greer PB. Reduction of the effect of non-uniform backscatter from an E-type support arm of a Varian a-Si EPID used for dosimetry. Phys Med Biol. 2010;55:6617-6632. https://doi.org/10.1088/0031-9155/55/22/003
  25. Boylan C, McWilliam A, Johnstone E, Rowbottom C. The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry. J Appl Clin Med Phys. 2012;13:4023. https://doi.org/10.1120/jacmp.v13i6.4023
  26. Bertelsen A, Lorenzen EL, Brink C. Validation of a new control system for Elekta accelerators facilitating continuously variable dose rate. Med Phys. 2011;38:4802-4810. https://doi.org/10.1118/1.3615621
  27. Nicolini G, Clivio A, Cozzi L, Fogliata A, Vanetti E. On the impact of dose rate variation upon RapidArc® implementation of volumetric modulated arc therapy. Med Phys. 2011;38:264-271. https://doi.org/10.1118/1.3528214
  28. Baily NA, Horn RA, Kampp TD. Fluoroscopic visualization of megavoltage therapeutic x ray beams. Int J Radiat Oncol Biol Phys. 1980;6:935-939. https://doi.org/10.1016/0360-3016(80)90341-7
  29. Boon SN, van Luijk P, Schippers JM, Meertens H, Denis JM, Vynckier S, et al. Fast 2D phantom dosimetry for scanning proton beams. Med Phys. 1998;25:464-475. https://doi.org/10.1118/1.598221
  30. Ding L, Wu Q, Wang Q, Li Y, Perks RM, Zhao L. Advances on inorganic scintillator-based optic fiber dosimeters. EJNMMI Phys. 2020;7:60. https://doi.org/10.1186/s40658-020-00327-6
  31. Goiffon V, Virmontois C, Magnan P, Cervantes P, Corbière F, Estribeau M, et al. Radiation damages in CMOS image sensors: testing and hardening challenges brought by deep sub-micrometer CIS processes. Paper presented at: Proc. SPIE 7826, Sensors, Systems, and Next-Generation Satellites XIV, 78261S; 2010 Sep 20-23; Toulouse, France.
  32. Tan J, Büttgen B, Theuwissen AJP. Radiation effects on CMOS image sensors due to X-Rays. Paper presented at: The Eighth International Conference on Advanced Semiconductor Devices and Microsystems; 2010 Oct 25-27; Smolenice, Slovakia. p. 279-282.
  33. Huber A, Sergienko G, Kinna D, Huber V, Milocco A, Mercadier L. Response of the imaging cameras to hard radiation during JET operation. Fusion Eng Des. 2017;123:669-673. https://doi.org/10.1016/j.fusengdes.2017.03.167
  34. Boon SN, van Luijk P, Böhringer T, Coray A, Lomax A, Pedroni E, et al. Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniques. Med Phys. 2000;27:2198-2208. https://doi.org/10.1118/1.1289372
  35. Petric MP, Robar JL, Clark BG. Development and characterization of a tissue equivalent plastic scintillator based dosimetry system. Med Phys. 2006;33:96-105. https://doi.org/10.1118/1.2140118
  36. Collomb-Patton V, Boher P, Leroux T, Fontbonne JM, Vela A, Batalla A. The DOSIMAP, a high spatial resolution tissue equivalent 2D dosimeter for LINAC QA and IMRT verification. Med Phys. 2009;36:317-328. https://doi.org/10.1118/1.3013703
  37. Guillot M, Beaulieu L, Archambault L, Beddar S, Gingras L. A new water-equivalent 2D plastic scintillation detectors array for the dosimetry of megavoltage energy photon beams in radiation therapy. Med Phys. 2011;38:6763-6774. https://doi.org/10.1118/1.3664007
  38. Zin HM, Harris EJ, Osmond JP, Allinson NM, Evans PM. Towards real-time VMAT verification using a prototype, high-speed CMOS active pixel sensor. Phys Med Biol. 2013;58:3359-3375. https://doi.org/10.1088/0031-9155/58/10/3359
  39. Miao X, Sim T. Ambient image recovery and rendering from flash photographs. Paper presented at: IEEE International Conference on Image Processing 2005; 2005 Sep 14; Genova, Italy. p. II-1038.
  40. Field R. Introduction to clinical digital photography. Gauteng: Modern Dentistry Media, 2017 [cited 2024 Sep 22]. Available from: http://www.moderndentistrymedia.com/dec_jan2018/field.pdf
  41. Wueller D. Evaluating digital cameras. Paper presented at: Proc. SPIE 6069, Digital Photography II, 60690K; 2006 Feb 15-19; San Jose, CA, USA.
  42. Hsu WF, Chuang KW, Hsu YC. Comparisons of the camera OECF, the ISO speed, and the SFR of digital still-picture cameras. Paper presented at: Proc. SPIE 4080, Input/Output and Imaging Technolgies II; 2000 Jul 26-28; Taipei, Taiwan.
  43. Hu Y, Zhu TC. Backscatter correction factor for megavoltage photon beam. Med Phys. 2011;38:5563-5568. https://doi.org/10.1118/1.3633903
  44. Zhu TC, Ahnesjö A, Lam KL, Li XA, Ma CM, Palta JR, et al.; AAPM Therapy Physics Committee Task Group 74. Report of AAPM Therapy Physics Committee Task Group 74: in-air output ratio, Sc, for megavoltage photon beams. Med Phys. 2009;36:5261-5291. https://doi.org/10.1118/1.3227367
  45. Hesamian MH, Jia W, He X, Kennedy P. Deep learning techniques for medical image segmentation: achievements and challenges. J Digit Imaging. 2019;32:582-596. https://doi.org/10.1007/s10278-019-00227-x
  46. Zhou T, Ruan S, Canu S. A review: deep learning for medical image segmentation using multi-modality fusion. Array. 2019;3-4:100004. https://doi.org/10.1016/j.array.2019.100004
  47. Ronneberger O, Fischer P, Brox T. U-Net: convolutional networks for biomedical image segmentation. Paper presented at: Medical Image Computing and Computer-Assisted Intervention – MICCAI 2015; 2015 Oct 5-9; Munich, Germany.
  48. Ye JC, Han Y, Cha E. Deep convolutional framelets: a general deep learning framework for inverse problems. SIAM J Imaging Sci. 2018;11:991-1048. https://doi.org/10.1137/17M1141771
  49. Han Y, Ye JC. Framing U-Net via deep convolutional framelets: application to sparse-view CT. IEEE Trans Med Imaging. 2018;37:1418-1429. https://doi.org/10.1109/TMI.2018.2823768
  50. Ge Y, Wei D, Xue Z, Wang Q, Zhou X, Zhan Y, et al. Unpaired Mr to CT Synthesis with explicit structural constrained adversarial learning. Paper presented at: 2019 IEEE 16th International Symposium on Biomedical Imaging (ISBI 2019); 2019 Apr 8-11; Venice, Italy. p. 1096-1099.
  51. Ge Y, Xue Z, Cao T, Liao S. Unpaired whole-body MR to CT synthesis with correlation coefficient constrained adversarial learning. Paper presented at: Proc. SPIE 10949, Medical Imaging 2019: Image Processing, 1094905; 2019 Mar 15; San Diego, CA, USA.
  52. Lee D, Jeong SW, Kim SJ, Cho H, Park W, Han Y. Improvement of megavoltage computed tomography image quality for adaptive helical tomotherapy using cycleGAN-based image synthesis with small datasets. Med Phys. 2021;48:5593-5610. https://doi.org/10.1002/mp.15182
  53. He K, Zhang X, Ren S, Sun J. Deep residual learning for image recognition. arXiv. 2015. https://doi.org/10.48550/arXiv.1512.03385
  54. Isola P, Zhu JY, Zhou T, Efros AA. Image-to-image translation with conditional adversarial networks. Paper presented at: 2017 IEEE Conference on Computer Vision and Pattern Recognition (CVPR); 2017 Jul 21-26; Honolulu, HI, USA.
  55. Lasio G, Guerrero M, Goetz W, Lima F, Baulch JE. Effect of varying dose-per-pulse and average dose rate in X-ray beam irradiation on cultured cell survival. Radiat Environ Biophys. 2014;53:671-676. https://doi.org/10.1007/s00411-014-0565-2
  56. Posar JA, Davis J, Brace O, Sellin P, Griffith MJ, Dhez O, et al. Characterization of a plastic dosimeter based on organic semiconductor photodiodes and scintillator. Phys Imaging Radiat Oncol. 2020;14:48-52. https://doi.org/10.1016/j.phro.2020.05.007
  57. Nyquist H. Certain factors affecting telegraph speed. Bell Syst Tech J. 1924;3:324-346. https://doi.org/10.1002/j.1538-7305.1924.tb01361.x
  58. Beck A, Teboulle M. A fast iterative shrinkage-thresholding algorithm for linear inverse problems. SIAM J Imaging Sci. 2009;2:183-202. https://doi.org/10.1137/080716542
  59. Martinet B. [Brève communication. Régularisat ion d'inéquations variationnelles par approximations successives]. Revue française d'informatique et de recherche opérationnelle. Série rouge. 1970;4:154-158. French. https://doi.org/10.1051/m2an/197004R301541
  60. Combettes PL, Wajs VR. Signal recovery by proximal forward-backward splitting. Multiscale Model Simul. 2005;4:1168-1200. https://doi.org/10.1137/050626090
  61. Siddon RL. Fast calculation of the exact radiological path for a three-dimensional CT array. Med Phys. 1985;12:252-255. https://doi.org/10.1118/1.595715
  62. Heil CE, Walnut DF. Continuous and discrete wavelet transforms. SIAM Rev. 1989;31:628-666. https://doi.org/10.1137/1031129
  63. Kingma DP, Ba J. Adam: a method for stochastic optimization. arXiv. 2017. https://doi.org/10.48550/arXiv.1412.6980
  64. Cheon W, Jung H, Lee M, Lee J, Kim SJ, Cho S, et al. Development of a time-resolved mirrorless scintillation detector. PLoS One. 2021;16:e0246742. https://doi.org/10.1371/journal.pone.0246742
  65. Levitan E, Herman GT. A maximum a posteriori probability expectation maximization algorithm for image reconstruction in emission tomography. IEEE Trans Med Imaging. 1987;6:185-192. https://doi.org/10.1109/TMI.1987.4307826
  66. Biguri A, Dosanjh M, Hancock S, Soleimani M. TIGRE: a MATLAB-GPU toolbox for CBCT image reconstruction. Biomed Phys Eng Express. 2016;2:055010. https://doi.org/10.1088/2057-1976/2/5/055010
  67. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656-661. https://doi.org/10.1118/1.598248
  68. Depuydt T, Van Esch A, Huyskens DP. A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation. Radiother Oncol. 2002;62:309-319. https://doi.org/10.1016/S0167-8140(01)00497-2
  69. Losasso T. IMRT delivery performance with a varian multileaf collimator. Int J Radiat Oncol Biol Phys. 2008;71:S85-S88. https://doi.org/10.1016/j.ijrobp.2007.06.082
  70. Li JG, Dempsey JF, Ding L, Liu C, Palta JR. Validation of dynamic MLC-controller log files using a two-dimensional diode array. Med Phys. 2003;30:799-805. https://doi.org/10.1118/1.1567951
  71. Zygmanski P, Kung JH, Jiang SB, Chin L. Dependence of f luence errors in dynamic IMRT on leaf-positional errors varying with time and leaf number. Med Phys. 2003;30:2736-2749. https://doi.org/10.1118/1.1598674
  72. Zeidan OA, Li JG, Ranade M, Stell AM, Dempsey JF. Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device. Med Phys. 2004;31:463-476. https://doi.org/10.1118/1.1644518
  73. Zeng GL. Maximum-likelihood expectation-maximization algorithm versus windowed filtered backprojection algorithm: a case study. J Nucl Med Technol. 2018;46:129-132. https://doi.org/10.2967/jnmt.117.196311
  74. Hughes J. Automated analysis of Varian log files for advanced radiotherapy treatment verification: a multicenter study [dissertation]. Crawley: University of Western Australia; 2015.
  75. Manikandan A, Sarkar B, Holla R, Vivek TR, Sujatha N. Quality assurance of dynamic parameters in volumetric modulated arc therapy. Br J Radiol. 2012;85:1002-1010. https://doi.org/10.1259/bjr/19152959
  76. Rangaraj D, Zhu M, Yang D, Palaniswaamy G, Yaddanapudi S, Wooten OH, et al. Catching errors with patient-specific pretreatment machine log file analysis. Pract Radiat Oncol. 2013;3:80-90. https://doi.org/10.1016/j.prro.2012.05.002
  77. Agnew A, Agnew CE, Grattan MW, Hounsell AR, McGarry CK. Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries. Phys Med Biol. 2014;59:N49-N63. https://doi.org/10.1088/0031-9155/59/9/N49
  78. Calvo-Ortega JF, Teke T, Moragues S, Pozo M, Casals-Farran J. A Varian DynaLog file-based procedure for patient dose-volume histogram-based IMRT QA. J Appl Clin Med Phys. 2014;15:4665. https://doi.org/10.1120/jacmp.v15i2.4665
  79. Bedford JL, Warrington AP. Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat Oncol Biol Phys. 2009;73:537-545. https://doi.org/10.1016/j.ijrobp.2008.08.055