Abstract
Purpose: The enhanced leaf modeling (ELM) algorithm, introduced in Eclipse v18, includes rounded leaf tip geometry to enhance the accuracy of multi-leaf collimator (MLC) modeling. This study aimed to assess the clinical consistency of empirically optimized dosimetric leaf gap (DLG) and transmission parameters used in the previous version (v16) when applied to the upgraded ELM algorithm. Methods: The DLG and transmission values optimized in Eclipse v16 were re-evaluated using the ELM model in v18. Because the underlying MLC modeling functions differ fundamentally between versions, the measured data in v18 were scaled to match the reference dose calculated in v16, allowing a direct comparison. Transmission factors and DLG values were compared between the two versions. Then, clinical plans, monitor unit (MU) calculations, and gamma passing rates (GPRs) were analyzed using portal dosimetry. Results: After scaling, the transmission factors between v16 and v18 demonstrated only minor differences of up to 0.0002, whereas the DLG values showed marked differences in both magnitude and sign. The allowable DLG range in v18 was between -0.1 cm and 0.05 cm, as defined by the manufacturer. Clinical plan comparisons revealed MU differences within ±1% for all beam conditions except those delivered on TrueBeam linear accelerator using 10 MV flattening-filter-free beams, which had a maximum deviation of -1.4%. All recalculated plans achieved GPRs above 90%, satisfying the clinical acceptance criteria. Conclusions: Despite fundamental changes in the MLC modeling algorithm, empirically optimized DLG and transmission values from the previous version remained clinically valid in the ELM-based modeling environment. These results demonstrate the robustness and reliability of the previously established parameters.