Validation of Clinically Optimized Multi-Leaf Collimator Parameters Using Enhanced Leaf Modeling in the AcurosXB Algorithm

  • Jin Jegal (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Yoonsuk Huh (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Inbum Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Bo-Wi Cheon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Euntaek Yoon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Chang Heon Choi (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jung-in Kim (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Seonghee Kang (Department of Radiation Oncology, Seoul National University Hospital)
  • Received : 2025.10.20
  • Accepted : 2025.11.13
  • Published : 2025.12.31

Abstract

Purpose: The enhanced leaf modeling (ELM) algorithm, introduced in Eclipse v18, includes rounded leaf tip geometry to enhance the accuracy of multi-leaf collimator (MLC) modeling. This study aimed to assess the clinical consistency of empirically optimized dosimetric leaf gap (DLG) and transmission parameters used in the previous version (v16) when applied to the upgraded ELM algorithm. Methods: The DLG and transmission values optimized in Eclipse v16 were re-evaluated using the ELM model in v18. Because the underlying MLC modeling functions differ fundamentally between versions, the measured data in v18 were scaled to match the reference dose calculated in v16, allowing a direct comparison. Transmission factors and DLG values were compared between the two versions. Then, clinical plans, monitor unit (MU) calculations, and gamma passing rates (GPRs) were analyzed using portal dosimetry. Results: After scaling, the transmission factors between v16 and v18 demonstrated only minor differences of up to 0.0002, whereas the DLG values showed marked differences in both magnitude and sign. The allowable DLG range in v18 was between -0.1 cm and 0.05 cm, as defined by the manufacturer. Clinical plan comparisons revealed MU differences within ±1% for all beam conditions except those delivered on TrueBeam linear accelerator using 10 MV flattening-filter-free beams, which had a maximum deviation of -1.4%. All recalculated plans achieved GPRs above 90%, satisfying the clinical acceptance criteria. Conclusions: Despite fundamental changes in the MLC modeling algorithm, empirically optimized DLG and transmission values from the previous version remained clinically valid in the ELM-based modeling environment. These results demonstrate the robustness and reliability of the previously established parameters.

Keywords

References

  1. Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84:967-996. https://doi.org/10.1259/bjr/22373346
  2. Cho B. Intensity-modulated radiation therapy: a review with a physics perspective. Radiat Oncol J. 2018;36:1-10. https://doi.org/10.3857/roj.2018.00122
  3. Saez J, Bar-Deroma R, Bogaert E, Cayez R, Chow T, Clark CH, et al. Universal evaluation of MLC models in treatment planning systems based on a common set of dynamic tests. Radiother Oncol. 2023;186:109775. https://doi.org/10.1016/j.radonc.2023.109775
  4. Paganini L, Reggiori G, Stravato A, Palumbo V, Mancosu P, Lobefalo F, et al. MLC parameters from static fields to VMAT plans: an evaluation in a RT-dedicated MC environment (PRIMO). Radiat Oncol. 2019;14:216. https://doi.org/10.1186/s13014-019-1421-y
  5. Van Esch A, Kulmala A, Rochford R, Kauppinen J, Harju A. Testing of an enhanced leaf model for improved dose calculation in a commercial treatment planning system. Med Phys. 2022;49:7754-7765. https://doi.org/10.1002/mp.v49.12
  6. Lin CY, Shiau AC, Ji JH, Lee CJ, Wang TH, Hsu SH, et al. A simple method for determining dosimetric leaf gap with cross-field dose width for rounded leaf-end multileaf collimator systems. Radiat Oncol. 2018;13:222. https://doi.org/10.1186/s13014-018-1164-1
  7. Vieillevigne L, Khamphan C, Saez J, Hernandez V. On the need for tuning the dosimetric leaf gap for stereotactic treatment plans in the Eclipse treatment planning system. J Appl Clin Med Phys. 2019;20:68-77. https://doi.org/10.1002/acm2.2019.20.issue-7
  8. Moktan H, Jiang H, Li HH, Guida K. Impact of enhanced leaf model on dose calculation accuracy in single-isocenter multitarget stereotactic radiosurgery treatments. J Appl Clin Med Phys. 2025;26:e70039. https://doi.org/10.1002/acm2.v26.6
  9. DiCostanzo DJ, Ayan AS. A novel methodology for the optimization of transmission and dosimetric leaf gap parameters. J Appl Clin Med Phys. 2022;23:e13565. https://doi.org/10.1002/acm2.v23.5
  10. Xue J, Wang H, Barbee D, Schmidt M, Das IJ. A practical method to optimize quality assurance results of arc therapy plans in beam modeling. J Med Phys. 2018;43:106-111. https://doi.org/10.4103/jmp.JMP_144_17
  11. Panagi R, Caines R, Rowbottom CG. Dosimetric sensitivity of an enhanced leaf model (ELM) for individual versus averaged machines. J Appl Clin Med Phys. 2025;26:e14621. https://doi.org/10.1002/acm2.v26.4
  12. Yang Y, Shoniyozov K, Li T, Lockamy V, Bieda M, Teo BK, et al. Evaluation of the accuracy of an enhanced MLC leaf model for Linac-based stereotactic radiosurgery. J Appl Clin Med Phys. 2025;26:e70143. https://doi.org/10.1002/acm2.v26.7
  13. Popple R, Cardan R, Sullivan R, Pogue JA, Awe B, Fiveash J. Effect of enhanced multileaf collimator leaf modeling on accuracy of dose calculation for single isocenter multiple target VMAT radiosurgery. Cureus 2025;17:a1408.