Clinical Validation of EPIbeam as a Patient-Specific Pre-Treatment Quality Assurance System

  • Euntaek Yoon (Interdisciplinary Program in Bioengineering, Graduate School, Seoul National University) ;
  • Inbum Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Yoonsuk Huh (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Bo-Wi Cheon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jin Jegal (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Hyeongmin Jin (Biomedical Research Institute, Seoul National University Hospital) ;
  • Sung Hyun Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Seonghee Kang (Biomedical Research Institute, Seoul National University Hospital) ;
  • Chang Heon Choi (Biomedical Research Institute, Seoul National University Hospital) ;
  • Jong Min Park (Biomedical Research Institute, Seoul National University Hospital) ;
  • Jung-in Kim (Biomedical Research Institute, Seoul National University Hospital) ;
  • Jaeman Son (Biomedical Research Institute, Seoul National University Hospital)
  • Received : 2025.10.30
  • Accepted : 2025.12.03
  • Published : 2025.12.31

Abstract

Purpose: This study aimed to clinically validate EPIbeam as a patient-specific pretreatment quality assurance (QA) system. Methods: EPIbeam performs QA by comparing the predicted dose from the intensity-modulated radiation therapy (IMRT) plan with the portal dose reconstructed from Elekta iViewGT images using a kernel-based algorithm. Verification included large square field tests (22×22-25×25 cm2) with 6- and 10-MV beams to assess field-size limitations, as well as a comparison of patient-specific QA results between EPIbeam and MapCHECK3 for 54 IMRT and volumetric modulated arc therapy plans across various treatment sites. Additional patient plans with extended field sizes were analyzed to further assess dose accuracy and clinical feasibility. Results: In the square field verification, all 6-MV fields up to 24×24 cm2 achieved gamma passing rates above 90% under the 3%/3 mm and 3%/2 mm criteria, whereas larger fields showed noticeable declines under stricter criteria. Similar patterns were observed for the 10-MV beams with slightly higher overall passing rates. Dose maps revealed underestimation at the four field corners. Patient-specific QA results were comparable to or higher than those from MapCHECK3. Extended-field plans maintained passing rates above 90% even under the strict 2%/1 mm criterion. Minor corner dose underestimations had a negligible clinical impact. Conclusions: EPIbeam demonstrated consistent and reliable performance for clinical plans within a 24×24 cm2 field size, with results comparable or superior to those of MapCHECK3. These results confirm the clinical validation of EPIbeam as a robust and reliable patient-specific pretreatment QA system.

Keywords

References

  1. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: recommendations of AAPM Task Group No. 218. Med Phys. 2018;45:e53-e83. https://doi.org/10.1002/mp.2018.45.issue-4
  2. Palta JR, Liu C, Li JG. Quality assurance of intensity-modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2008; 71:S108-S112. https://doi.org/10.1016/j.ijrobp.2007.05.092
  3. Jomy J, Sharma R, Lu R, Chen D, Ataalla P, Kaushal S, et al. Clinical impact of radiotherapy quality assurance results in contemporary cancer trials: a systematic review and meta-analysis. Radiother Oncol. 2025;207:110875. https://doi.org/10.1016/j.radonc.2025.110875
  4. Dogan N, Mijnheer BJ, Padgett K, Nalichowski A, Wu C, Nyflot MJ, et al. AAPM Task Group Report 307: use of EPIDs for patient-specific IMRT and VMAT QA. Med Phys. 2023; 50:e865-e903.
  5. Baek SH, Choi SH, Han MJ, Cho GS, Jang W, Kim JS, et al. Clinical efficacy of an electronic portal imaging device versus a physical phantom tool for patient-specific quality assurance. Life (Basel). 2022;12:1923. https://doi.org/10.3390/life12111923
  6. Kim Y, Park H, Ryu H, Oh DH, Chun M. Comparison of portal dosimetry and 3D diode arrays for volumetric modulated arc therapy patient-specific quality assurance. Prog Med Phys. 2025;36:45-51. https://doi.org/10.14316/pmp.2025.36.3.45
  7. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25:656-661. https://doi.org/10.1118/1.598248
  8. Muralidhar KR, Nishaanth B, Kumar AV, Srinivas P, Kolluru VRR, Kumar R, et al. Clinical application of portal dosimetry for pre-treatment dosimetric verification across various diagnoses using electronic portal imaging integrated with Epibeam. Am J Clin Exp Med. 2025;13:1-7. https://doi.org/10.11648/j.ajcem
  9. Ahnesjö A, Saxner M, Trepp A. A pencil beam model for photon dose calculation. Med Phys. 1992;19:263-273. https://doi.org/10.1118/1.596856
  10. Nijsten SM, van Elmpt WJ, Jacobs M, Mijnheer BJ, Dekker AL, Lambin P, et al. A global calibration model for a-Si EPIDs used for transit dosimetry. Med Phys. 2007;34:3872-3884. https://doi.org/10.1118/1.2776244
  11. Seneclauze A, Boissard P, Baudier T, Dupuis P, Malet C. 9 MLC error detection with EPIBeam for VMAT patient quality assurance. Phys Medica. 2018;56:6. https://doi.org/10.1016/j.ejmp.2018.09.022
  12. Ouakkad A, Goubert M, Vieillevigne L, Husson F, Parent L. EP-1724 Delivery error sensitivity of an EPID based pretreatment control for FFF dynamic arc therapy. Radiother Oncol. 2019;133:S929. https://doi.org/10.1016/S0167-8140(19)32144-9
  13. Lewis D, Micke A, Yu X, Chan MF. An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan. Med Phys. 2012;39:6339-6350. https://doi.org/10.1118/1.4754797
  14. Palmer AL, Nash D. Radiochromic film dosimetry in radiotherapy: a survey of current practice in the United Kingdom. Br J Radiol. 2024;97:646-651. https://doi.org/10.1093/bjr/tqae008
  15. Altaf S, Iqbal K, Akram M, Buzdar SA. Quality assurance of intensity-modulated radiotherapy treatment planning: a dosimetric comparison. J Radiother Pract. 2018;17:396-402. https://doi.org/10.1017/S1460396918000195
  16. Son J, Baek T, Lee B, Shin D, Park SY, Park J, et al. A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy. Radiol Oncol. 2015; 49:307-313. https://doi.org/10.1515/raon-2015-0021