Comparative Evaluation of a Multi-Purpose Winston-Lutz Test Phantom with Lock Bar Fixture for Linear Accelerator Quality Assurance

  • Jihae Lee (Department of Radiation Oncology, Ewha Womans University Mokdong Hospital) ;
  • Young Jin Won (Department of Radiation Oncology, Uijeongbu Eulji Medical Center, Eulji University) ;
  • Sang Hyoun Choi (Department of Radiation Oncology, Korea Institute of Radiological and Medical Sciences) ;
  • Samju Cho (Department of Radiation Oncology, Ewha Womans University Mokdong Hospital)
  • Received : 2025.11.14
  • Accepted : 2025.12.01
  • Published : 2025.12.31

Abstract

Purpose: To evaluate the performance of a multi-purpose Winston-Lutz (W/L) test phantom device with a lock bar fixture compared to commercially available phantoms for verifying linear accelerator isocenter accuracy. Methods: W/L tests were performed using three phantom systems. Measurements included multiple imaging protocols (Ewha Winston-Lutz: 8 beams, Sun Nuclear Winston-Lutz [SWL] and Sun Nuclear comprehensive Winston-Lutz: 12 beams each), as well as gantry rotation tests to analyze mechanical isocenter accuracy. Field sizes of 1×1 cm2 and 2×2 cm2 were compared, and dose dependency was evaluated using 2 monitor unit (MU) versus 20 MU portal imaging. Analyses were performed using both PipsPro (DoseLab Pro) and SunMachine (Sun Patient) software platforms. Statistical significance was assessed using independent samples t-tests (P<0.05) and Bland-Altman analysis to evaluate inter-software agreement. Results: The developed W/L test phantom demonstrated superior measurement accuracy, with mean 3D position errors of 0.65-0.74 mm across all protocols compared to the commercial W/L phantom using the SWL protocol. Measurement uncertainty for the EP phantom was quantified as ±0.22 mm (k=1), demonstrating that observed differences in measurement precision exceed the expected variability. The novel phantom exhibited lower standard deviations (0.16-0.24 mm) compared to the commercial phantoms. Both analysis software platforms produced consistent results with excellent agreement. Conclusions: The novel multi-purpose W/L phantom with a lock bar fixture provides superior measurement precision and consistent performance across different imaging protocols and analysis platforms, supporting its suitability for routine quality assurance in modern radiation therapy.

Keywords

References

  1. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197-4212. https://doi.org/10.1118/1.3190392
  2. International Electrotechnical Commission (IEC). IEC 62083:2009. Medical electrical equipment - requirements for the safety of radiotherapy treatment planning systems. IEC. 2009;62083.
  3. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. 2010;37:4078-4101. https://doi.org/10.1118/1.3438081
  4. Solberg TD, Medin PM, Mullins J, Li S. Quality assurance of immobilization and target localization systems for frameless stereotactic cranial and extracranial hypofractionated radiotherapy. Int J Radiat Oncol Biol Phys. 2008;71:S131-S135. https://doi.org/10.1016/j.ijrobp.2007.05.097
  5. Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int J Radiat Oncol Biol Phys. 1988;14:373-381. https://doi.org/10.1016/0360-3016(88)90446-4
  6. Rowshanfarzad P, Sabet M, O'Connor DJ, Greer PB. Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques. J Appl Clin Med Phys. 2011;12:3645. https://doi.org/10.1120/jacmp.v12i4.3645
  7. Treuer H, Hoevels M, Luyken K, Gierich A, Kocher M, Müller RP, et al. On isocentre adjustment and quality control in linear accelerator based radiosurgery with circular collimators and room lasers. Phys Med Biol. 2000;45:2331-2342. https://doi.org/10.1088/0031-9155/45/8/318
  8. Ohira S, Ueda Y, Akino Y, Hashimoto M, Masaoka A, Hirata T, et al. HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach. Radiat Oncol. 2018;13:13. https://doi.org/10.1186/s13014-017-0948-z
  9. Bissonnette JP, Balter PA, Dong L, Langen KM, Lovelock DM, Miften M, et al. Quality assurance for image-guided radiation therapy utilizing CT-based technologies: a report of the AAPM TG-179. Med Phys. 2012;39:1946-1963. https://doi.org/10.1118/1.3690466