Abstract
Purpose: To evaluate the performance of a multi-purpose Winston-Lutz (W/L) test phantom device with a lock bar fixture compared to commercially available phantoms for verifying linear accelerator isocenter accuracy. Methods: W/L tests were performed using three phantom systems. Measurements included multiple imaging protocols (Ewha Winston-Lutz: 8 beams, Sun Nuclear Winston-Lutz [SWL] and Sun Nuclear comprehensive Winston-Lutz: 12 beams each), as well as gantry rotation tests to analyze mechanical isocenter accuracy. Field sizes of 1×1 cm2 and 2×2 cm2 were compared, and dose dependency was evaluated using 2 monitor unit (MU) versus 20 MU portal imaging. Analyses were performed using both PipsPro (DoseLab Pro) and SunMachine (Sun Patient) software platforms. Statistical significance was assessed using independent samples t-tests (P<0.05) and Bland-Altman analysis to evaluate inter-software agreement. Results: The developed W/L test phantom demonstrated superior measurement accuracy, with mean 3D position errors of 0.65-0.74 mm across all protocols compared to the commercial W/L phantom using the SWL protocol. Measurement uncertainty for the EP phantom was quantified as ±0.22 mm (k=1), demonstrating that observed differences in measurement precision exceed the expected variability. The novel phantom exhibited lower standard deviations (0.16-0.24 mm) compared to the commercial phantoms. Both analysis software platforms produced consistent results with excellent agreement. Conclusions: The novel multi-purpose W/L phantom with a lock bar fixture provides superior measurement precision and consistent performance across different imaging protocols and analysis platforms, supporting its suitability for routine quality assurance in modern radiation therapy.