Approximate Modeling of the Zap-XⓇ Linear Accelerator Head Using mcnpxTM

  • Kyoung Won Jang (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Heuijin Lim (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Manwoo Lee (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Jung Ki Kim (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Kyo Chul Shin (Department of Radiation Oncology, Dankook University Hospital) ;
  • In-Ho Jung (Department of Neurosurgery, Dankook University Hospital) ;
  • Sang Koo Lee (Department of Neurosurgery, Dankook University Hospital) ;
  • Ki-Hwan Kim (Department of Radiation Oncology, Chungnam National University Hospital) ;
  • Dong Hyeok Jeong (Research Center, Dongnam Institute of Radiological and Medical Sciences)
  • Received : 2025.11.02
  • Accepted : 2025.12.01
  • Published : 2025.12.31

Abstract

Modeling of the radiosurgery system is required to study dosimetry, dose calculation, and radiation shielding for small radiosurgical fields. In this study, a 3 MV linear accelerator head of the Zap-XⓇ radiosurgery system was modeled using a Monte Carlo code, and the target thickness and photon beam spot size consistent with the measured dose distribution were determined. The mcnpxTM (Monte Carlo N-particle eXtended v2.4.0) code was used in the geometry modeling and dose calculations in a water phantom. To determine the correct modeling parameters, the photon energy spectrum and depth dose were calculated for various target thicknesses, and the beam profile was calculated for different photon beam spot sizes and compared with the measured values. The components of the Zap-XⓇ head were modeled, and the target thickness, along with the photon beam spot size, was determined based on measured dose distributions. The calculated percentage depth dose values closely matched the measured values when the target thickness ranged from 1.8 to 2.4 mm. Optimal agreement between calculated and measured beam profiles was observed when the photon beam spot exhibited a full-width at half maximum of 0.8 mm. Modeling parameters for the ZAP-XⓇ radiosurgery system head were identified that matched well with the measured data. The established model of the Zap-XⓇ system will be widely used in radiosurgical beam research.

Keywords

Acknowledgement

This work was supported by the Dongnam Institute of Radiological and Medical Sciences (DIRAMS) grant funded by the Korea government (MSIT) under grant 50493-2025.

References

  1. Chen T, Riley P, Chung W, Zhou L, Tao H, Guo J. Design, fabrication and verification of a 3MeV S-band medical linear accelerator. Paper presented at: 14th International Particle Accelerator Conference; 2023 May 7-12; Venice, Italy.
  2. Weidlich GA, Bodduluri M, Achkire Y, Lee C, Adler JR Jr. Characterization of a Novel 3 megavolt linear accelerator for dedicated intracranial stereotactic radiosurgery. Cureus. 2019;11:e4275. https://doi.org/10.7759/cureus.4275
  3. Weidlich GA, Chung W, Kolli S, Thirunarayanan I, Loysel T. Characterization of the ZAP-X® peripheral dose fall-off. Cureus. 2021;13:e13972. https://doi.org/10.7759/cureus.13972
  4. Weidlich GA, Schneider MB, Simcic V, Oostman Z, Adler JR Jr. Self-shielding for the ZAP-X®: revised characterization and evaluation. Cureus. 2021;13:e13660. https://doi.org/10.7759/cureus.13660
  5. Andreo P. A comparison between calculated and experimental kQ photon beam quality correction factors. Phys Med Biol. 2000;45:L25-L38. https://doi.org/10.1088/0031-9155/45/9/101
  6. Zink K, Wulff J. Monte Carlo calculations of beam quality correction factors kQ for electron dosimetry with a parallel-plate Roos chamber. Phys Med Biol. 2008;53:1595-1607. https://doi.org/10.1088/0031-9155/53/6/006
  7. International Atomic Energy Agency (IAEA). Dosimetry of small static fields used in external beam radiotherapy. Technical reports series No. 483. IAEA. 2017;483.
  8. Lu S, Deng H. Monte Carlo modeling and verification of 6 MV linear accelerator. World J Eng Technol. 2022;10:213-223. https://doi.org/10.4236/wjet.2022.102012
  9. Gray T, Bassiri N, Kirby N, Stathakis S, Mayer KM. Implementation of a simple clinical linear accelerator beam model in MCNP6 and comparison with measured beam characteristics. Appl Radiat Isot. 2020;155:108925. https://doi.org/10.1016/j.apradiso.2019.108925
  10. Rogers DW, Faddegon BA, Ding GX, Ma CM, We J, Mackie TR. BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 1995;22:503-524. https://doi.org/10.1118/1.597552
  11. Waters LS. MCNPXTM user's manual. Version 2.3.0. Los Alamos National Laboratory. 2002; LA-UR-02-2607.