Feasibility Study of a 4D Respiratory Phantom for Quality Assurance of Particle Therapy

  • Received : 2025.11.10
  • Accepted : 2025.11.28
  • Published : 2025.12.31

Abstract

Purpose: Respiratory motion in particle therapy necessitates quality assurance (QA) tools capable of decoupling mechanical/temporal performance from complex dosimetric perturbations. This study aims to develop a four-dimensional internal-external respiratory motion phantom (4D IERM Phantom) to simulate patient-specific respiratory patterns and tumor motion for particle therapy QA applications. Methods: The 4D IERM Phantom consists of a control unit and an operating unit. The control unit was constructed using a printed circuit board based on an Arduino Mega 2560, while the operating unit was divided into two main parts: the abdominal and chest regions. The abdominal region mimics 3D target motion in the anterior-posterior (AP), superior-inferior, and left-right directions using three stepper motors, whereas the chest region simulates 1D motion in the AP direction with a single stepper motor. Additionally, the 4D IERM Phantom is designed to accommodate dose measurements using a multi-active volume ion chamber and a MatriXX for both the abdominal and chest regions. For validation, actual patient respiratory data with a 50-ms time interval and a 0.2 mm step size were used. Results: The 4D IERM Phantom was validated using respiratory data from 72 actual patients. The phantom successfully simulated patient-specific respiratory patterns and tumor motion across all four independent axes. Motion fidelity analysis demonstrated high accuracy: the Pearson correlation coefficient (r) exceeded 0.97 across all four axes. Quantitative metrics confirmed high precision, with root mean square error values ranging from 0.1008 mm to 0.7162 mm, and maximum absolute error contained within approximately 2 mm for the AP axes. Conclusions: The 4D IERM Phantom mimics complex, patient-specific respiratory motion and provides a robust, feasible solution for the isolated mechanical and temporal validation of 4D motion management systems in particle therapy.

Keywords

Acknowledgement

This work was supported by grants from the National Cancer Center of Korea (NCC-2410960-2).

References

  1. Korreman SS. Motion in radiotherapy: photon therapy. Phys Med Biol. 2012;57:R161-R191.
  2. Lin L, Kang M, Huang S, Mayer R, Thomas A, Solberg TD, et al. Beam-specific planning target volumes incorporating 4D CT for pencil beam scanning proton therapy of thoracic tumors. J Appl Clin Med Phys. 2015;16:5678.
  3. Keall PJ, Mageras GS, Balter JM, Emery RS, Forster KM, Jiang SB, et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys. 2006;33:3874-3900.
  4. Shirato H, Seppenwoolde Y, Kitamura K, Onimura R, Shimizu S. Intrafractional tumor motion: lung and liver. Semin Radiat Oncol. 2004;14:10-18.
  5. Doyen J, Falk AT, Floquet V, Hérault J, Hannoun-Lévi JM. Proton beams in cancer treatments: clinical outcomes and dosimetric comparisons with photon therapy. Cancer Treat Rev. 2016;43:104-112.
  6. Rietzel E, Bert C. Respiratory motion management in particle therapy. Med Phys. 2010;37:449-460.
  7. Giraud P, Morvan E, Claude L, Mornex F, Le Pechoux C, Bachaud JM, et al. Respiratory gating techniques for optimization of lung cancer radiotherapy. J Thorac Oncol. 2011;6:2058-2068.
  8. Bert C, Grözinger SO, Rietzel E. Quantification of interplay effects of scanned particle beams and moving targets. Phys Med Biol. 2008;53:2253-2265.
  9. Bert C, Durante M. Motion in radiotherapy: particle therapy. Phys Med Biol. 2011;56:R113-R144.
  10. Boria AJ, Uh J, Pirlepesov F, Stuckey JC, Axente M, Gargone MA, et al. Interplay effect of target motion and pencil-beam scanning in proton therapy for pediatric patients. Int J Part Ther. 2018;5:1-10.
  11. Ball HJ, Santanam L, Senan S, Tanyi JA, van Herk M, Keall PJ. Results from the AAPM Task Group 324 respiratory motion management in radiation oncology survey. J Appl Clin Med Phys. 2022;23:e13810.
  12. Li H, Dong L, Bert C, Chang J, Flampouri S, Jee KW, et al. AAPM Task Group Report 290: respiratory motion management for particle therapy. Med Phys. 2022;49:e50-e81.
  13. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, et al. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36:4197-4212.
  14. Woods K, Rong Y. Technical report: TG-142 compliant and comprehensive quality assurance tests for respiratory gating. Med Phys. 2015;42:6488-6497.
  15. Dunn L, Kron T, Johnston PN, McDermott LN, Taylor ML, Callahan J, et al. A programmable motion phantom for quality assurance of motion management in radiotherapy. Australas Phys Eng Sci Med. 2012;35:93-100.
  16. Acquah GF, Kyeremeh PO, Hasford F, Boadu M, Sosu EK, Inkoom S. Evaluation of metallic implant artifact on photon beam calculation algorithms using a CIRS thorax phantom. J Radiat Res Appl Sci. 2018;11:347-349.
  17. Chang J, Suh TS, Lee DS. Development of a deformable lung phantom for the evaluation of deformable registration. J Appl Clin Med Phys. 2010;11:3081.
  18. Kim J, Lee Y, Shin H, Ji S, Park S, Kim J, et al. Development of deformable moving lung phantom to simulate respiratory motion in radiotherapy. Med Dosim. 2016;41:113-117.
  19. Black DG, Yazdi YO, Wong J, Fedrigo R, Uribe C, Kadrmas DJ, et al. Design of an anthropomorphic PET phantom with elastic lungs and respiration modeling. Med Phys. 2021;48:4205-4217.
  20. Mukumoto N, Nakamura M, Yamada M, Takahashi K, Akimoto M, Miyabe Y, et al. Development of a four-axis moving phantom for patient-specific QA of surrogate signal-based tracking IMRT. Med Phys. 2016;43:6364.
  21. Ionascu D, Jiang SB, Nishioka S, Shirato H, Berbeco RI. Internal-external correlation investigations of respiratory induced motion of lung tumors. Med Phys. 2007;34:3893-3903.
  22. Wiersma RD, McCabe BP, Belcher AH, Jensen PJ, Smith B, Aydogan B. Technical note: high temporal resolution characterization of gating response time. Med Phys. 2016;43:2802-2806.
  23. Liu Z, Pan L, Ma T, Lu HM, Wang Y. Comprehensive beam delivery latency evaluation for gated proton therapy system using customized multi-channel signal acquisition platform. J Appl Clin Med Phys. 2024;25:e14349.
  24. Giżyńska MK, Seppenwoolde Y, Jm Heijmen B. Tumor tracking with non-linear internal/external correlation models in the presence of respiratory motion baseline drifts and phase shifts. Phys Med. 2024;118:103295.
  25. Chen H, Zhong Z, Yang Y, Chen J, Zhou L, Zhen X, et al. Internal motion estimation by internaI-external motion modeling for lung cancer radiotherapy. Sci Rep. 2018;8:3677.
  26. Aranjo B, Soori PK, Talukder P. Stepper motor drives for robotic applications. Paper presented at: 2012 IEEE International Power Engineering and Optimization Conference Melaka, Malaysia; 2012 Jun 6-7; Melaka, Malaysia.
  27. Boye D, Lomax T, Knopf A. Mapping motion from 4D-MRI to 3D-CT for use in 4D dose calculations: a technical feasibility study. Med Phys. 2013;40:061702.
  28. Krieger M, Klimpki G, Fattori G, Hrbacek J, Oxley D, Safai S, et al. Experimental validation of a deforming grid 4D dose calculation for PBS proton therapy. Phys Med Biol. 2018;63:055005.
  29. Serban M, Heath E, Stroian G, Collins DL, Seuntjens J. A deformable phantom for 4D radiotherapy verification: design and image registration evaluation. Med Phys. 2008;35:1094-1102.
  30. Shin DS, Kang SH, Kim DS, Kim TH, Kim KH, Koo HJ, et al. Development of an advanced deformable phantom to analyze dose differences due to respiratory motion. Prog Med Phys. 2017;28:1-10.