Clinical Performance of FractionLab Using Log Files for Patient-Specific Quality Assurance in Stereotactic Radiosurgery Using Truebeam Equipped with HyperArc

  • Ji Young Bae (Department of Physics, Yeungnam University) ;
  • Sung Yeop Kim (Department of Physics, Yeungnam University) ;
  • Jae Won Park (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Jaehyeon Park (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Ji Woon Yea (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Yoon Young Jo (Department of Radiation Oncology, Yeungnam University Medical Center) ;
  • Se An Oh (Department of Radiation Oncology, Yeungnam University Medical Center)
  • Received : 2025.01.17
  • Accepted : 2025.03.20
  • Published : 2025.06.30

Abstract

Purpose: This study evaluates the clinical performance of FractionLab, a log file-based quality assurance (QA) tool for stereotactic radiosurgery (SRS) using the TrueBeam system with HyperArc technology. It compares FractionLab's effectiveness and accuracy with traditional portal dose image prediction (PDIP) methods to determine its suitability for patient-specific QA in advanced radiotherapy techniques. Methods: A retrospective analysis was conducted on 43 patients who underwent SRS equipped with HyperArc for brain metastases or small intracranial tumors. QA was performed using both FractionLab and PDIP. Gamma-passing rates under various criteria (3%/3 mm, 2%/1 mm) were calculated and statistically analyzed using paired t-tests. The comparison focused on FractionLab's ability to detect discrepancies between planned and delivered doses. Results: FractionLab demonstrated comparable or superior performance to PDIP across most criteria, particularly at 3%/3 mm, where it achieved higher gamma-passing rates under less stringent conditions. However, under stricter criteria, such as 2%/1 mm and 0.1/0.1, FractionLab's gamma-passing rate was significantly lower, highlighting the need for further optimization in these scenarios. Statistical analysis confirmed significant differences in the passing rates under specific conditions. This emphasizes FractionLab's potential for efficiency and accuracy in HyperArc QA. Conclusions: FractionLab is considered a reliable and efficient QA tool for HyperArc treatments, offering several advantages in terms of operational efficiency and precision. Its performance aligns with traditional methods under standard conditions; however, optimization is needed for stricter criteria. Furthermore, the findings support integrating FractionLab into clinical workflows for advanced radiotherapy QA.

Keywords

Acknowledgement

This work was supported by a grant from the Chunma medical research foundation, Korea, 2024.

References

  1. Wong FHC, Moleme PA, Ali OA, Mugabe KV. Clinical implementation of HyperArc. Phys Eng Sci Med. 2022;45:577-587. https://doi.org/10.1007/s13246-022-01123-5
  2. Kamst O, Desai P. Evaluation of HyperArcTM using film and portal dosimetry quality assurance. Phys Eng Sci Med. 2023;46:57-66. https://doi.org/10.1007/s13246-022-01197-1
  3. Ohira S, Ueda Y, Akino Y, Hashimoto M, Masaoka A, Hirata T, et al. HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach. Radiat Oncol. 2018;13:13. https://doi.org/10.1186/s13014-017-0948-z
  4. Pokhrel D, Bernard ME, Johnson J, Fabian D, Kudrimoti M. HyperArc VMAT stereotactic radiotherapy for locally recurrent previously-irradiated head and neck cancers: plan quality, treatment delivery accuracy, and efficiency. J Appl Clin Med Phys. 2022;23:e13561. https://doi.org/10.1002/acm2.13561
  5. Kim J, Kim TG, Park B, Kim H, Song YG, Lee HW, et al. Dosimetric comparison between RapidArc and HyperArc in hippocampal-sparing whole-brain radiotherapy with a simultaneous integrated boost. Med Dosim. 2024;49:69-76. https://doi.org/10.1016/j.meddos.2023.08.007
  6. Raza GH, Capone L, Tini P, Giraffa M, Gentile P, Minniti G. Single-isocenter multiple-target stereotactic radiosurgery for multiple brain metastases: dosimetric evaluation of two automated treatment planning systems. Radiat Oncol. 2022;17:116. https://doi.org/10.1186/s13014-022-02086-3
  7. Pan M, Xu W, Sun L, Wang C, Dong S, Guan Y, et al. Dosimetric quality of HyperArc in boost radiotherapy for single glioblastoma: comparison with CyberKnife and manual VMAT. Radiat Oncol. 2023;18:8. https://doi.org/10.1186/s13014-022-02150-y
  8. Park SY, Choi N, Jang NY. Dosimetric evaluations of HyperArc and RapidArc in stereotactic radiosurgery for a single brain metastasis. Prog Med Phys. 2024;35:36-44. https://doi.org/10.14316/pmp.2024.35.2.36
  9. Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994;21:581-618. https://doi.org/10.1118/1.597316
  10. Popple RA, Brown MH, Thomas EM, Willey CD, Cardan RA, Covington EL, et al. Transition from manual to automated planning and delivery of volumetric modulated arc therapy stereotactic radiosurgery: clinical, dosimetric, and quality assurance results. Pract Radiat Oncol. 2021;11:e163-e171. https://doi.org/10.1016/j.prro.2020.10.013
  11. Agnew A, Agnew CE, Grattan MW, Hounsell AR, McGarry CK. Monitoring daily MLC positional errors using trajectory log files and EPID measurements for IMRT and VMAT deliveries. Phys Med Biol. 2014;59:N49-N63. https://doi.org/10.1088/0031-9155/59/9/N49
  12. Chan LT, Tan YI, Tan PW, Leong YF, Khor JS, Teh MW, et al. Comparing log file to measurement-based patient-specific quality assurance. Phys Eng Sci Med. 2023;46:303-311. https://doi.org/10.1007/s13246-023-01219-6
  13. Oh SA, Kim SY, Park J, Park JW, Yea JW. Clinical performance of FractionLab in patient-specific quality assurance for intensity-modulated radiotherapy: a retrospective study. J Yeungnam Med Sci. 2022;39:108-115. https://doi.org/10.12701/yujm.2021.01123
  14. Chow VUY, Kan MWK, Chan ATC. Patient-specific quality assurance using machine log files analysis for stereotactic body radiation therapy (SBRT). J Appl Clin Med Phys. 2020; 21:179-187. https://doi.org/10.1002/acm2.13053
  15. Vanderstraeten B, Reynaert N, Paelinck L, Madani I, De Wagter C, De Gersem W, et al. Accuracy of patient dose calculation for lung IMRT: a comparison of Monte Carlo, convolution/superposition, and pencil beam computations. Med Phys. 2006;33:3149-3158. https://doi.org/10.1118/1.2241992
  16. Vazquez-Quino LA, Huerta-Hernandez CI, Rangaraj D. Clinical experience with machine log file software for volumetric-modulated arc therapy techniques. Proc (Bayl Univ Med Cent). 2017;30:276-279. https://doi.org/10.1080/08998280.2017.11929614
  17. Rasch D, Kubinger KD, Moder K. The two-sample t test: pre-testing its assumptions does not pay off. Stat Pap (Berl). 2011;52:219-231. https://doi.org/10.1007/s00362-009-0224-x
  18. Pasler M, Kaas J, Perik T, Geuze J, Dreindl R, Künzler T, et al. Linking log files with dosimetric accuracy--a multiinstitutional study on quality assurance of volumetric modulated arc therapy. Radiother Oncol. 2015;117:407-411. https://doi.org/10.1016/j.radonc.2015.11.005
  19. Lew KS, Chua CGA, Koh CWY, Lee JCL, Park SY, Tan HQ. Prediction of portal dosimetry quality assurance results using log files-derived errors and machine learning techniques. Front Oncol. 2022;12:1096838. https://doi.org/10.3389/fonc.2022.1096838
  20. Kerns JR, Childress N, Kry SF. A multi-institution evaluation of MLC log files and performance in IMRT delivery. Radiat Oncol. 2014;9:176. https://doi.org/10.1186/1748-717X-9-176
  21. Kim TK. Understanding one-way ANOVA using conceptual figures. Korean J Anesthesiol. 2017;70:22-26. https://doi.org/10.4097/kjae.2017.70.1.22
  22. Zhang J, Cheng Z, Fan Z, Zhang Q, Zhang X, Yang R, et al. A feasibility study for in vivo treatment verification of IMRT using Monte Carlo dose calculation and deep learningbased modelling of EPID detector response. Radiat Oncol. 2022;17:31. https://doi.org/10.1186/s13014-022-01999-3