Abstract
Purpose: This study evaluated the clinical feasibility and dosimetric accuracy of a newly implemented ceiling-mounted spoiler (CMS) system for total body irradiation (TBI) and established beam parameters for extended source-to-surface distance (SSD) setups. Methods: The CMS consisted of modular acrylic panels (1.5 cm thick, 50×147 cm2) mounted on ceiling rails aligned with the treatment position, which was located 400 cm from the isocenter. Detachable magnetic sections enabled full-body coverage and compact storage. Uniformity was assessed using a Farmer-type ionization chamber to compare dose variations between section centers and bridge regions. Surface and midline doses were measured with and without a spoiler, as well as with CMS, using metal-oxide-semiconductor field-effect transistor detectors. Tissue maximum ratios (TMRs) were verified against treatment planning system (TPS) values at 100 and 400 cm SSD. Spoiler, tray, and compensator factors were measured at a source-to-detector distance of 400 cm and used for calculating the monitor unit. Dose validation was performed on an anthropomorphic phantom at five anatomical sites. Results: The CMS exhibited <0.1% dose variation between section regions, ensuring uniform scatter. Surface dose enhancement was comparable to a conventional spoiler and superior to no spoiler. TMRs matched TPS values within ~1.5% at 100 cm SSD. However, the deviations were greater at 400 cm SSD, especially for 6 MV. Phantom measurements revealed dose deviations within ±7.5%, with larger discrepancies observed in the shoulder and head regions. Conclusions: The CMS provides effective surface dose enhancement and uniform scattering as well as improved mechanical usability, supporting its clinical application in extended SSD TBI treatments.