Clinical Commissioning of a Ceiling-Mounted Spoiler for Total Body Irradiation

  • Jin Jegal (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Yoonsuk Huh (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Inbum Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Bo-Wi Cheon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Euntaek Yoon (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Seonghee Kang (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jaeman Son (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Sung Hyun Lee (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Chang Heon Choi (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Jung-in Kim (Department of Radiation Oncology, Seoul National University Hospital) ;
  • Hyeongmin Jin (Department of Radiation Oncology, Seoul National University Hospital)
  • Received : 2025.08.08
  • Accepted : 2025.09.12
  • Published : 2025.09.30

Abstract

Purpose: This study evaluated the clinical feasibility and dosimetric accuracy of a newly implemented ceiling-mounted spoiler (CMS) system for total body irradiation (TBI) and established beam parameters for extended source-to-surface distance (SSD) setups. Methods: The CMS consisted of modular acrylic panels (1.5 cm thick, 50×147 cm2) mounted on ceiling rails aligned with the treatment position, which was located 400 cm from the isocenter. Detachable magnetic sections enabled full-body coverage and compact storage. Uniformity was assessed using a Farmer-type ionization chamber to compare dose variations between section centers and bridge regions. Surface and midline doses were measured with and without a spoiler, as well as with CMS, using metal-oxide-semiconductor field-effect transistor detectors. Tissue maximum ratios (TMRs) were verified against treatment planning system (TPS) values at 100 and 400 cm SSD. Spoiler, tray, and compensator factors were measured at a source-to-detector distance of 400 cm and used for calculating the monitor unit. Dose validation was performed on an anthropomorphic phantom at five anatomical sites. Results: The CMS exhibited <0.1% dose variation between section regions, ensuring uniform scatter. Surface dose enhancement was comparable to a conventional spoiler and superior to no spoiler. TMRs matched TPS values within ~1.5% at 100 cm SSD. However, the deviations were greater at 400 cm SSD, especially for 6 MV. Phantom measurements revealed dose deviations within ±7.5%, with larger discrepancies observed in the shoulder and head regions. Conclusions: The CMS provides effective surface dose enhancement and uniform scattering as well as improved mechanical usability, supporting its clinical application in extended SSD TBI treatments.

Keywords

References

  1. Wills C, Cherian S, Yousef J, Wang K, Mackley HB. Total body irradiation: a practical review. Appl Rad Oncol. 2016;(2):11-17.
  2. Paix A, Antoni D, Waissi W, Ledoux MP, Bilger K, Fornecker L, et al. Total body irradiation in allogeneic bone marrow transplantation conditioning regimens: a review. Crit Rev Oncol Hematol. 2018;123:138-148. https://doi.org/10.1016/j.critrevonc.2018.01.011
  3. Kim MH, Rhu SY, Lim DS, Song JY. Effects of low-dose fractionated total body irradiation on murine immune system. J Radiat Prot. 2014;39:134-141. https://doi.org/10.14407/jrp.2014.39.3.134
  4. Hoseinnezhad Zarghani E, Geraily G, Haddad P, Esfahani M, Farzin M, Rastjoo A, et al. Dosimetric comparison of AP/PA and bilateral geometries for total body irradiation treatment. Radiat Environ Biophys. 2021;60:663-672. https://doi.org/10.1007/s00411-021-00933-7
  5. Patel RP, Warry AJ, Eaton DJ, Collis CH, Rosenberg I. In vivo dosimetry for total body irradiation: five-year results and technique comparison. J Appl Clin Med Phys. 2014;15:4939. https://doi.org/10.1120/jacmp.v15i4.4939
  6. Pearlman R, Hanna R, Burmeister J, Abrams J, Dominello M. Adverse effects of total body irradiation: a two-decade, single institution analysis. Adv Radiat Oncol. 2021;6:100723. https://doi.org/10.1016/j.adro.2021.100723
  7. Carruthers SA, Wallington MM. Total body irradiation and pneumonitis risk: a review of outcomes. Br J Cancer. 2004;90:2080-2084. https://doi.org/10.1038/sj.bjc.6601751
  8. Khan FM. The physics of radiation therapy. 4th ed. Lippincott Williams & Wilkins; 2009.
  9. Ravichandran R, Binukumar JP, Davis CA, Al Rahbi Z, Balakrishnan R, Al Mandhari Z. Total body irradiation (TBI): preliminary experience on clinical implementation. J Med Phys. 2013;38:210-211. https://doi.org/10.4103/0971-6203.121200
  10. Nelson G, Sarkar V, Szegedi M, Molineu A, Olch AJ, Kunz JN, et al. Validation of Acuros for total body irradiation at extended distance. J Appl Clin Med Phys. 2024;25:e14468.
  11. Hoseinnezhad E, Geraily G, Esfahani M, Farzin M, Gholami S. Comparison of calculated and measured basic dosimetric parameters for total body irradiation with 6- and 18-MV photon beams. J Radiother Pract. 2021;20:66-70. https://doi.org/10.1017/S1460396919001067
  12. Chairman JVD, Galvin JM, Glasgow GP, Podgorsak EB. The physical aspects of total and half body photon irradiation. AAPM Report. 1986; 17.
  13. Choi CH, Park JM, Park SY, Chun M, Han JH, Cho JD, et al. Prediction of midline dose from entrance and exit dose using OSLD measurements for total body irradiation. J Radiat Prot Res. 2017;42:77-82. https://doi.org/10.14407/jrpr.2017.42.2.77
  14. Akbas U, Donmez Kesen N, Koksal C, Bilge H. Surface and buildup region dose measurements with markus parallel plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for high-energy photon beams. Adv High Energy Phys. 2016;2016:8361028. https://doi.org/10.1155/2016/8361028