Comparison of Portal Dosimetry and 3D Diode Arrays for Volumetric Modulated Arc Therapy Patient-Specific Quality Assurance

  • Yoohyeon Kim (Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital) ;
  • Hyojun Park (Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital) ;
  • Hyejo Ryu (Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital) ;
  • Do Hoon Oh (Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital) ;
  • Minsoo Chun (Department of Radiation Oncology, Chung-Ang University Gwangmyeong Hospital)
  • Received : 2025.07.21
  • Accepted : 2025.09.04
  • Published : 2025.09.30

Abstract

Purpose: This study evaluated the clinical performance of an electronic portal imaging device-based portal dosimetry system (EPIbeam) in comparison with 3D diode array systems (ArcCHECK® and SRS MapCHECK®) for patient-specific quality assurance (PSQA) in volumetric modulated arc therapy (VMAT). Methods: Thirty-six VMAT treatment plans, each comprising two arc fields, were analyzed. For PSQA, gamma analysis was performed using ArcCHECK® for standard VMAT plans with a 2 mm/2% criterion and SRS MapCHECK® for stereotactic VMAT plans with a 1 mm/2% criterion. EPIbeam was evaluated using the same gamma criteria for both plan types. Results: Commissioning of EPIbeam demonstrated excellent agreement, with gamma passing rates (GPRs) exceeding 99.5% for all fields and energies using the 2 mm/2% criterion. For PSQA, the standard VMAT plans exhibited significantly higher GPRs with EPIbeam (98.7%±1.3%) than with ArcCHECK® (95.9%±2.8%) (P<0.05). In contrast, for the stereotactic VMAT plans, SRS MapCHECK® outperformed EPIbeam, with GPRs of 95.7%±2.6% and 88.9%±7.5%, respectively (P<0.05). Conclusions: EPIbeam demonstrated high accuracy and efficiency for standard VMAT PSQA, suggesting its suitability as a standalone tool. However, for stereotactic VMAT plans involving small targets and high dose rates, additional QA using high-resolution detectors is recommended to ensure robust verification.

Keywords

References

  1. Palma D, Vollans E, James K, Nakano S, Moiseenko V, Shaffer R, et al. Volumetric modulated arc therapy for delivery of prostate radiotherapy: comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy. Int J Radiat Oncol Biol Phys. 2008;72:996-1001. https://doi.org/10.1016/j.ijrobp.2008.02.047
  2. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: recommendations of AAPM Task Group No. 218. Med Phys. 2018;45:e53-e83. https://doi.org/10.1002/mp.2018.45.issue-4
  3. Dogan N, Mijnheer BJ, Padgett K, Nalichowski A, Wu C, Nyflot MJ, et al. AAPM Task Group Report 307: use of EPIDs for patient-specific IMRT and VMAT QA. Med Phys. 2023;50:e865-e903. https://doi.org/10.1002/mp.v50.8
  4. Min S, Choi YE, Kwak J, Cho B. Practical approach for pretreatment verification of IMRT with flattening filter free (FFF) beams using Varian Portal Dosimetry. J Appl Clin Med Phys. 2014;16:4934.
  5. Baek SH, Choi SH, Han MJ, Cho GS, Jang W, Kim JS, et al. Clinical efficacy of an electronic portal imaging device versus a physical phantom tool for patient-specific quality assurance. Life (Basel). 2022;12:1923. https://doi.org/10.3390/life12111923
  6. Muralidhar KR, Nishaanth B, Kumar AV, Srinivas P, Kolluru VRR, Kumar R, et al. Clinical application of portal dosimetry for pre-treatment dosimetric verification across various diagnoses using electronic portal imaging integrated with epibeam. Am J Clin Exp Med. 2025;13:1-7. https://doi.org/10.11648/j.ajcem
  7. Chaswal V, Weldon M, Gupta N, Chakravarti A, Rong Y. Commissioning and comprehensive evaluation of the Arc-CHECK cylindrical diode array for VMAT pretreatment delivery QA. J Appl Clin Med Phys. 2014;15:4832. https://doi.org/10.1120/jacmp.v15i4.4832
  8. Xu Q, Huynh K, Nie W, Rose MS, Chawla AK, Choe KS, et al. Implementing and evaluating a high-resolution diode array for patient-specific quality assurance of robotic brain stereotactic radiosurgery/radiotherapy. J Appl Clin Med Phys. 2022;23:e13569. https://doi.org/10.1002/acm2.v23.5
  9. Kim JI, Park SY, Kim HJ, Kim JH, Ye SJ, Park JM. The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol. 2014;9:167. https://doi.org/10.1186/1748-717X-9-167
  10. Manna S, Gayen S, Varshney S. Analysis of non-coplanar dose reconstruction for evaluating pre-treatment quality assurance in stereotactic treatments. Radiat Phys Chem. 2025;234:112790. https://doi.org/10.1016/j.radphyschem.2025.112790
  11. Price RA Jr, Veltchev I, Lin T, Eldib A, Chen L, Jin L, et al. Evaluating suggested stricter gamma criteria for linac-based patient-specific delivery QA in the conventional and SBRT environments. Phys Med. 2022;100:72-80. https://doi.org/10.1016/j.ejmp.2022.06.005
  12. Chun M, An HJ, Kwon O, Oh DH, Park JM, Kim JI. Impact of plan parameters and modulation indices on patient-specific QA results for standard and stereotactic VMAT. Phys Med. 2019;62:83-94. https://doi.org/10.1016/j.ejmp.2019.05.005
  13. Oh DH, Choi JH, Ryu H, Chun M. Quantification of treatment plan deliverability in breast volumetric-modulated arc therapy with agility multi-leaf collimator. In Vivo. 2024;38:2254-2260. https://doi.org/10.21873/invivo.13690
  14. Pardo E, Novais JC, Molina López MY, Ruiz Maqueda S. On flattening filter-free portal dosimetry. J Appl Clin Med Phys. 2016;17:132-145. https://doi.org/10.1120/jacmp.v17i4.6147
  15. Chendi A, Botti A, Orlandi M, Sghedoni R, Iori M, Cagni E. EPID-based 3D dosimetry for pre-treatment FFF VMAT stereotactic body radiotherapy plan verification using dosimetry CheckTM. Phys Med. 2021;81:227-236. https://doi.org/10.1016/j.ejmp.2020.12.014